jakobheyman / xdripJH

Modified version of xDrip+ (Libre-2 OOP2 setup)

Geek Repo:Geek Repo

Github PK Tool:Github PK Tool

xdripJH

Modified version of xDrip+

xDrip+ source: https://github.com/NightscoutFoundation/xDrip

Major notes

  • Modifications done primarily for Libre-2 OOP2 setup
  • Display 1 min data instead of original 5 min
  • Default calibration changed including optional historic recalculation of bg values (see below)
  • Use OOP raw or calibrated bg as main data (Less common settings > Other misc. options)
  • Use raw/calibrated data as blood glucose values if there is no calibration available
  • Display raw data (Graph Settings > Display raw data plot)
  • Display OOP calibrated data (Graph Settings > Display OOP calibrated data plot)
  • Display last OOP calibrated value in home screen (Graph Settings > Display OOP calibrated bg value)
  • Optionally modify Libre-2 raw values with multiplication and/or +/- (Advanced settings for Libre 2)
  • Display mmol glucose values with two decimal digits
  • Extended potential min and max blood glucose values
  • 1 second X (time) resolution in graph
  • Keep Y range constant in main graph and optionally modify min and max Y (Graph Settings)
  • Modify X/Y display when switching between portrait/landscape (Graph Settings)
  • Horizontal zoom in main graph and vertical zoom in preview graph
  • Optionally set custom vibration pattern for low/high alerts (example: 0,150,70,150,70,500)
  • Optionally hide noise line (Graph Settings)
  • Calibration Graph uses the same unit (mg/dl or mmol/l) for raw and calibrated values
  • Modified basal TBR and micro bolus icons display
  • Statistics buttons added (DD) with custom start (D1) and stop (D2) dates
  • There may be issues/problems introduced by the modifications
  • There is no proper testing of the application

Calibration notes

  • Slope and intercept calculated with weights decreasing linearly with time from calibration point
  • The weighting can be controlled with these three parameters (Advanced Calibration)
    • Calibration weight days - time over which the weight decreases from 1 to 0
    • Calibration weight days initial - weight time used at the start of a new sensor
    • Calibration weight days initial transition - transition time from initial to standard weight days
  • Use raw data as blood glucose if there is no calibration available (no initial warm up period)
  • No specific initial calibration used (slope = 1 after the first calibration)
  • Optionally recalculate all affected calibrations and glucose values at calibration (Graph Settings)

About

Modified version of xDrip+ (Libre-2 OOP2 setup)

License:GNU General Public License v3.0


Languages

Language:Java 99.9%Language:Python 0.1%Language:Kotlin 0.0%Language:Shell 0.0%